Metrohm IC Net 2.0 User Manual
Page 29
![background image](/manuals/586518/29/background.png)
4.3 System functions
IC Net 2.0
21
SYSTEM / Control / Disconnect from workplace
Disconnect selected system from the workplace. If a system is
disconnected, manual control is not available for this system, but
all other system settings can be modified and saved.
Start deter-
mination
and
Startup hardware
is only possible if no other system
from the same system folder is connected. In this case, the sys-
tem is automatically connected.
4.3.2
Start/stop hardware and record baseline
SYSTEM / Control / Startup hardware (Measure Baseline)
With this menu item the following actions are executed automati-
cally:
•
Send
System startup values
to the devices.
•
Start Metrohm solvent delivery systems and Bischoff solvent
delivery systems.
•
Start Metrohm additional modules and Bischoff additional
modules.
•
Start Bischoff detectors.
•
Switch Metrohm Autosamplers and Bischoff Autosamplers to
remote control.
•
Start recording of the measurement signal using the method of
the connected system. Independently of the set chromatogram
Duration
, the measurement signal is recorded until the data
acquisition is stopped if a new determination is started with
Start determination
or by clicking the icon of the chroma-
togram window. In this case the user is asked if the recorded
baseline should be saved or not.
SYSTEM / Control / Shutdown hardware
With this menu item the devices are stopped as follows:
•
Stop Metrohm solvent delivery systems and Bischoff solvent
delivery systems.
•
Stop Metrohm additional modules and Bischoff additional
modulesVariotherm.
•
Stop Bischoff detectors.
•
Switch Metrohm Autosamplers and Bischoff Autosamplers to
local control.
•
Stop running determination.
•
Stop active sample queue.
SYSTEM / Control / Stop data acquisition
Stop recording of the baseline.