7.4 Integration

IC Net 2.0

115

If an integration event is added or modified, the

Edit integration

event

window is opened where the following parameters can be

set:

Time

Start time for integration event in min.

Event

List box with integration events to be se-

lected (see below).

Value

Parameter value for selected integration

event. This field appears for events that

demand input of an additional parameter.

Integration events for peak detection

Disable detection

Set this mode to stop detection of new

peaks. If a peak is available at the moment

of the event it is either finished (downslope

peaks) or rejected (upslope peaks).

Enable detection

Clear

Disable detection

mode.

Enable negative peaks

Enable detection of negative peaks. In

some cases this mode may result in insta-

bility of detection algorithm.

Disable negative peaks

Disable detection of negative peaks. This

event does not influence negative peaks

which already started.

Disable peak reject

Set this mode, if a small peak should not

be rejected because of its flat apex.

Enable peak reject

Clear

Disable peak reject