Metrohm IC Net 2.0 User Manual
Page 53
5.4 761 Compact IC
IC Net 2.0
45
Second column
Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the
Remote configura-
tion
tab.
Third column
Parameter for program instruction (see
below).
ENABLED
Enable program start (a disabled program
is not started).
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction
Parameter entry
Meaning
Valve
Inject, Fill
Switch injection valve to "INJECT"
or "FILL" position.
FullScale
50, 250, 1000
µµS/cm
Set full scale range to the se-
lected value.
ICPump
on, off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5 mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 25.0 MPa
Set maximum pressure limit for
the high-pressure pump to the de-
sired value.
Pmin
0.0 ... 25.0 MPa
Set minimum pressure limit for
the high-pressure pump to the de-
sired value.
Remote
0, 1, *, p
Set remote output lines 1...8 to
the desired values. For entry of the
first value, enter
1
,
0
,
p
or
*
. For en-
try of the other values, move the
cursor in front of the value to be
changed and enter
1
,
0
,
p
or
*
.