Metrohm IC Net 2.0 User Manual
Page 179

10.2 Perform batch reprocessing
IC Net 2.0
171
Use method from file for reprocessing
Selection of the desired chromatogram file
whose method should be used for reproc-
essing.
If the reprocessing includes reintegration
and/or recalibration, only chromatograms
recorded with the same method should
be loaded into the batch reprocessing
queue.
Open the chromatogram selected above in
the
Use method...
field.
Open all chromatograms of the batch re-
processing queue.
Open the batch reprocessing editor pro-
gram for editing the batch reprocessing ta-
ble (see section 10.3).
Reprocess sample runs
Reprocess all sample chromatograms
(
calibration level = 0
).
Reprocess calibration runs
Reprocess all calibration chromatograms
(
calibration level > 0
).
Update method file in
Save the method file
*.mtw
after reprocess-
ing if the method is changed.
Reintegrate
Reintegrate chromatograms according to
the current settings of the integration pa-
rameters and integration events.
Open the
Integration parameters
window for
modification of integration parameters and
integration events.
Recalibrate
Reprocess all calibration chromatograms (if
Reprocess calibration runs
is switched on)
and apply new calibration to all sample
chromatograms (if
Reprocess sample runs
is
switched on) by updating the concentration
table.
Default scheme
Default setting for recalibration reprocess-
ing. The two options
Apply final calibration...
and
Forget calibration points...
are switched
on. A new calibration is performed with the
calibration runs and the resulting new cali-