4.5 System state window
IC Net 2.0
27
Running
Running program or chromatogram data
acquisition.
Running program
(xxx min left)
Running time program with time display.
Waiting for INJECT
Waiting for "INJECT" to start program
and/or chromatogram data acquisition as
defined in the Start mode.
INJECT done
Injection valve has been switched to the
"INJECT" position.
Stopping
Determination has been stopped.
Finished
Program or chromatogram data acquisition
has been finished.
SHUTDOWN
System is shutdown.
4.5.2
General error messages
Detection of hard-
ware failed
Bad connection between PC and device or
device switched off (check connecting ca-
ble or switch on instrument).
4.5.3
Error messages for 761 Compact IC
LEAK DETECTED
The leak detector has detected a leak
(check IC system and connections).
E1
Program checksum wrong (call Metrohm
service).
E2
RAM faulty (call Metrohm service).
E200
Invalid instrument adjustment (call Metrohm
service).
E237
Storage of configuration values failed
(repeat last action; if error reappears, call
Metrohm service).
E238
Storage of instrument number failed (repeat
last action; if error reappears, call Metrohm
service).
E240
EEPROM faulty (call Metrohm service).
E258
Storage of setup values failed (repeat last
action; if error reappears, call Metrohm
service).