Metrohm IC Net 2.0 User Manual
Page 57

5.4 761 Compact IC
IC Net 2.0
49
Cell constant
Cell constant of the conductivity cell for
correct display of the absolute conductivity.
Enter the cell constant printed on the de-
tector block into this field.
Range:
0.1
...
1000 /cm
For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, ob-
serve the displayed conductivity and
change the cell constant until the correct
conductivity value is displayed.
Possibility for changing the break times for
injection valve
Valve
and suppressor mod-
ule
Suppressor
.
These values should only be changed af-
ter consulting the Metrohm Service.
Outputs
The
Outputs
tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event
Events for automatic remote signal output:
Inject
Switching of the injection valve to the
"INJECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Suppressor step
Switching of the suppressor module to the
next position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.