Metrohm IC Net 2.0 User Manual
Page 152

7 Methods
IC Net 2.0
144
type
The types of components are designated
by a one-letter code:
R
Reference component (used for peak iden-
tification).
S
Standard component.
C
Calibration standard in the case it is differ-
ent from the quantification standard.
?
Some points of corresponding peak are out
of ADC or detector range – suspicious re-
sult.
!
Component concentration is outside of
minimum (
min C
) and maximum limits
(
max C
), set in the component table.
p
Special component (it has one of the spe-
cial flags checked on in the component
window).
N
The component response is outside of
calibrated region.
In the case of peaks additional letters pre-
cede the component type information, indi-
cating, how peaks are separated from other
peaks, for example:
BD_
Peak that starts on the baseline (
B
) and
ends on the drop line (
D
) that separates it
from another adjacent peak.
BBR
Special case of a rider (
R
) peak that is tan-
gentially separated from the main one. A
main peak will have a third letter
H
(horse)
in this case.
The complete component type may look
like:
BBD : !R
.
group
Number of the group for the component.
spectral ratio
Ratio of responses on different channels for
multi-channel chromatograms.
name
Name of the component.
file name
File name of chromatogram. This column is
very useful for the purpose of processing of
exported data.
ident
Title (identifier) of the chromatogram. This
column is useful for the processing of ex-
ported data.