Metrohm IC Net 2.0 User Manual
Page 54
![background image](/manuals/586518/54/background.png)
5 Interfaces
IC Net 2.0
46
Program
END, RESET
The
END
flag can be used to end a
program, especially if the program
time should be longer than the
chromatogram duration. Additional
steps after this flag are not allowed.
The
RESET
flag is used to reset the
parameters to the system startup
values.
Suppressor
Switch suppressor module to the
next position.
Peristaltic
on, off
Switch on or off the peristaltic
pump.
Remote configuration
On the
Remote configuration
subtab user-defined remote com-
mands can be defined, which can be inserted into a time pro-
gram after being activated with
.
Name (1
st
column)
User-definable name of the remote com-
mand (e.g.
Start_766
).
Remote command (2
nd
column)
Setting the remote output lines 1
…
8.
0
(line off, inactive, open)
1
(line on, active, 0 V)
p
(pulse)
*
(leave line in current status)
For entry of the first value, enter
1
,
0
,
p
or
*
.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter
1
,
0
,
p
or
*
.
Add new remote command.
Delete selected remote command.
Activate the defined remote commands for
insertion into the time program.