Metrohm IC Net 2.0 User Manual
Page 40

5 Interfaces
IC Net 2.0
32
5.2.4
762 IC Interface window
762 icon / Open
The
762 IC Interface
window for parameter settings is opened by
selecting this menu option with the right mouse button or by
clicking the
762 icon
on the toolbar. It consists of three tabs
Init
,
RS settings
, and
Links
.
Init
The
Init
tab of the
762 IC Interface
window contains data acquisi-
tion parameters for the 762 IC Interface.
Label
Optional label to name the interface with
maximum 8 characters.
ROM version
Instrument program version number.
Measuring per second
Number of data points measured per s.
Entry range:
10, 20, 30, 50, 60 points/s
Range
Range for AD converter.
Example: converts
±
2500 mV to 2
24
bits
Entry range:
2500, 1250, 625, 312.5, 156.25,
78.125, 39.062 mV
Second press on RUN/STOP means 'stop' when measuring
If this option is enabled, the running data
acquisition will be stopped if the
RUN/STOP button on the 762 IC Interface
front panel is pressed. This is equal to
Stop
determination
.
If this option is disabled, the running data
acquisition will not be stopped if the
RUN/STOP button on the 762 IC Interface
front panel is pressed.