5 Interfaces

IC Net 2.0

50

Hardware error

Hardware error detected at the 761 Com-

pact IC (high-pressure pump, injection

valve, or suppressor not working correctly).

Alarm leak detector

Leak detector has detected solvent in the

instruments interior.

Remote line

Set remote output lines 1

8.

Selection:

0

(line off, inactive, open)

1

(line on, active, 0 V)

p

(pulse output)

*

(leave line in current status)

PowerOn values

Display of the power on startup values for

remote output lines set on the

Hardware

tab.

Pulse length

Length of a pulse in ms.

Alarm stops

The

Alarm stops

tab of the hardware settings window defines the

events for which the 761 Compact IC is stopped immediately. At

an alarm stop, high-pressure pump and peristaltic pump are

stopped immediately, the running determination and the active

sample queue are also stopped.

Event

Events for alarm stop:

Pump stop

Pump stopped because pressure limits are

exceeded.

Hardware error

Hardware error detected at the 761 Com-

pact IC (high-pressure pump, injection

valve, or suppressor not working correctly).

Leak

Leak detector has detected solvent in the

instruments interior. This information is also

stored in the instrument itself, so that it is

stopped automatically even without con-

nection to the PC.

Power on

Temporary power failure at the 761 Com-

pact IC hardware.