9.2 Sample queue control
IC Net 2.0
165
Amount
Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
C
s
= C
c
•
Amount
s
/
Amount
c
Shut down system after the queue finishes
If this option is enabled, the system hard-
ware is shut down after the sample queue
has been finished (all pumps are switched
off), but it remains connected.
Start execution of the sample queue from
the first line with
Started = 0
.
Stop execution of the sample queue after
the current determination has been fin-
ished.
Open the sample queue editor program for
editing the sample queue table (see section
9.3).
Close the sample queue overview table.
9.2.2
Start sample queue
SAMPLE QUEUE OVERVIEW /
Control / Start
Start execution of the sample queue from the first row with
Started
= 0
. For each row, the sample-specific data are transferred to the
method of the current determination and the parameter
Started
is
set to
1
for this row.
A running sample queue can be interrupted with
. It can
be restarted with
(with a running chromatogram only when
data acquisition is finished).
9.2.3
Pause sample queue
SAMPLE QUEUE OVERVIEW /
Control / Pause
Interrupt execution of the sample queue after the current determi-
nation has been finished.
A running sample queue can be interrupted with
. When
the current determination has been finished, the sample queue
can be restarted with
or stopped with
Control / Stop deter-
mination
in the
SYSTEM
window.