7 Methods

IC Net 2.0

118

Force horizontal

Cancel horizontal

base back

base back

Force horizontal base back

Set horizontal baseline for a single peak.

The baseline is drawn backwards from the

peak end point. The intersection of the

baseline with the signal is defined as peak

start point.

Cancel horizontal base back

Clear

Force horizontal base back

mode.

Enable baseline

Disable baseline

penetration

penetration

Enable baseline penetration

Allow crossing of the signal by the base-

line.

Disable baseline penetration

Clear

Enable baseline penetration

mode.

Integration events for integration parameters

Set width

Sets new

Width

parameter that supersedes

default linear growth of expected peak.

Set slope

Sets new

Slope

value.

Set min height

Sets new

Minimum peak height

value.

Set rider ratio

Sets new value of

Rider ratio

parameter.