3 batch reprocessing queue editor, 1 open batch reprocessing queue editor window, 2 batch reprocessing queue editor functions – Metrohm IC Net 2.0 User Manual
Page 183: 3 print batch reprocessing queue, 4 close batch reprocessing queue editor

10.3 Batch reprocessing queue editor
IC Net 2.0
175
10.3
Batch reprocessing queue editor
10.3.1 Open batch reprocessing queue editor window
REPROCESS /
This menu item opens the batch reprocessing editor program for
editing the batch reprocessing table.
No
Row number.
File name
Name of the chromatogram file (read-only).
Method
Name of the method file
*.mtw
used for re-
cording the chromatogram (read-only).
Ident
User defined identifier for chromatogram.
Will be placed into appropriate passport
field when starting the chromatogram.
Vial
Autosampler vial position to take sample
from. Will be placed into appropriate pass-
port field.
Volume
Injected volume of sample in
µ
L. Will be
placed into appropriate passport field.
Dilution
Sample dilution prior to injection. Will be
placed into appropriate passport field.
Amount
Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
C
s
= C
c
•
Amount
s
/
Amount
c
Will be placed into appropriate passport field.
Internal standard amount
Concentration of the internal standard
component for relative concentration cal-
culations. Will be placed into appropriate
passport field.
Calibration level
Calibration level (see section 7.5) for the
sample: Level
0
stays for normal analysis,
levels
1
and greater for calibration runs.
Correct filling of this column enables to cal-
culate calibration coefficients automatically.