Metrohm IC Net 2.0 User Manual

Page 51

background image

5.4 761 Compact IC

IC Net 2.0

43

IC pump

High-pressure pump.

Start high-pressure pump.

Stop high-pressure pump.

Peristaltic pump

Peristaltic pump.

Start peristaltic pump.

Stop peristaltic pump.

System startup values

The system startup values are sent and

applied to the 761 Compact IC each time

the system is connected, a determination is

started, or the values are sent manually

with

.

Flow, mL/min

Startup value for flow rate of the high-

pressure pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit

for high-pressure pump. This limit is con-

trolled even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0

MPa

Full scale,

µµS/cm

Startup value for full scale range.

Selection:

50

,

250

,

1000

µµS/cm

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1

Initial

The

Initial

tab of the

761 Compact IC

window for settings of system

startup values is only available for a disconnected system.

Flow, mL/min

Startup value for flow rate of the high-

pressure pump.

Entry range:

0.20

...

2.50 mL/min