Metrohm IC Net 2.0 User Manual
Page 51

5.4 761 Compact IC
IC Net 2.0
43
IC pump
High-pressure pump.
Start high-pressure pump.
Stop high-pressure pump.
Peristaltic pump
Peristaltic pump.
Start peristaltic pump.
Stop peristaltic pump.
System startup values
The system startup values are sent and
applied to the 761 Compact IC each time
the system is connected, a determination is
started, or the values are sent manually
with
.
Flow, mL/min
Startup value for flow rate of the high-
pressure pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit
for high-pressure pump. This limit is con-
trolled even without connection to the PC.
Entry range:
0.0
...
25.0 MPa
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:
0.0
...
25.0
MPa
Full scale,
µµS/cm
Startup value for full scale range.
Selection:
50
,
250
,
1000
µµS/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1
Initial
The
Initial
tab of the
761 Compact IC
window for settings of system
startup values is only available for a disconnected system.
Flow, mL/min
Startup value for flow rate of the high-
pressure pump.
Entry range:
0.20
...
2.50 mL/min