Metrohm IC Net 2.0 User Manual
Page 43

5.2 762 IC Interface
IC Net 2.0
35
Time program
On the
Program
tab of the
762 System [#]
window a user-defined
time program for the event output lines of the 762 IC Interface
can be entered. This program is started automatically as defined
in the
Start mode
window (see section 4.4.3) either at the moment
the determination is started (
Start with determination
) or at the
moment the sample is injected (
Start with inject
).
The
Program
tab contains the two following subpages:
Program
Main time program with all program steps.
Events configuration
Possibility for creation of user-defined re-
mote commands.
Program
On the
Program
subpage, program steps including time, program
instruction and parameter can be entered.
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last in-
struction with time entry.
Second column
Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the
Events configura-
tion
tab.
Third column
Parameter for program instruction (see
below).
ENABLED
Enable program start (a disabled program
is not started).