8 Programming

76

8.4.6 Plotarea (zoom)

Defines the range of the input signal to be
recorded on the chart. The complete range
or an enlarged section can be recorded. In-
put is in percent. The low value designates
the point of the measurement range which
becomes the new zero. The high value des-
ignates the new end value. The entire
measurement range (scaling start to scaling
end) corresponds to 100%.

The low value is calculated as follows:

The high value is calculated as follows:

The difference between low and
high value must be at least 10%.

Display

Edit

Selection/input

with keys

Enter

Continue with

PLOTAREA

E

2 PLOTAREA

Select recording channel

h

,

R

E

2

2

2 LOW: 010 %

Input start of
presentation range of
measurement value
(in percent of the
measurement range)

h

,

R

,

l

,

r

E

Check:
0

start

90%

3

3

2 HIGH: 100 %

Input end of presentation
range of measurement
value (in percent of the
measurement range)

h

,

R

,

l

,

r

E

Check:
end

100%

and end–start

10%

h

forwards

R

backwards

(smallest value to be recorded – start of scaling) /
(end of scaling – start of scaling)

h 100%

(largest value to be recorded – start of scaling) /
(end of scaling – start of scaling)

h 100%