beautypg.com

2 error messages, Error messages – JUMO 706021 LOGOLINE 500d Line Recorder Operating Manual User Manual

Page 125

background image

12 Appendix

125

12.2 Error messages

All error messages are displayed flashing
on the matrix display at regular intervals.
The other instrument functions remain unaf-
fected as far as posssible.

Display

Cause / Remedy

Status message

NO PAPER!

The chart cassette has been removed or the end of the
chart has been reached; a fresh chart has to be fitted.
 Section 9.2 ff
Caution: Insert the chart cassette by applying light up-
wards pressure

The take-up roll has not properly engaged with the
chart cassette
 Section 9.2.1

Error or fault on recorder

LOW BATTERY!

The battery for the real time clock and for RAM back-
up is discharged.

Please contact the nearest offfice or the factory.

ADJUST TIME!

This errror message is produced when the storage ca-
pacitor (Code 021,  Section 2.2) is discharged too far
during supply failure.

Event, supply failure and operating hour counters and
the system clock are reset.

The time must be set.

 Section 8.3.2

RELAY-ERROR!

The communication with the external relay module
ER8 is faulty.

Check the connection between the recorder and the
external relay module ER8.