8.2.4 Level inhibit and code request

The parameter level is protected by a 4-digit
codenumber.

Two codenumbers are available:

- codenumber for a limited set of parameters

(factory-set to 9200)

- codenumber for a full set of parameters

(factory-set to 9210)

Which parameters belong to the limited set
of parameters is defined in the set-up pro-
gram under Edit Instrument operation.

Parameters which are not required for the
applications or which are only programmed
once can be marked there accordingly. Af-
ter the new set-up has been transferred to
the pen recorder, they are not indicated any
more on the instrument, .

If one of the two codenumbers is entered,
the following are interrupted:

- signal acquisition, and

- recording

Additionally:

- events are no longer acquired,

- reports are aborted and reset,

- limit values are no longer monitored, and

- the outputs remain in their latest status.

The parameters can now be programmed.