8 Programming

55

8.3 Parameter level

At the parameter level it is possible to con-
figure general parameters.

From the operating level, the parameter lev-
el is reached by simultaneously pressing
keys

h

+

R

, from the configuration level 1

by pressing the

e

key.

The parameter level covers the following pa-
rameters:

- language

- date and time

- summer time

- display brightness

- relay limit settings

- dispaly time

If an incorrect codenumber is en-
tered, the recorder remains in nor-
mal operation. Signal acquisition,
processing and recording remain
activated.

The parameters can be viewed but
can not be programmed.

Following entry of a correct
codenumber the normal operation
is interrupted.

There is no signal acquisition,
processing or recording.

The relays of the external relay
module ER8 remain in their current
status. They are only operated
again when the recorder is at the
operating level or in the basic sta-
tus.

Display

Edit

Selection/input

with keys

Enter

Continue with

P-LEVEL

h

forwards

R

backwards