12 Appendix

127

12.3 Hardware fault

In case of one of the following faults the re-
cording is aborted and the error message
flashes on the display.

The relays are operated as for probe break

1

.

The recorder does not react to any event
and can not be operated.
Please contact the nearest office or the factory.

1. The action on probe break is configured

using the setup program or the parameter
relay output (

Section 8.6.1).

Display

Cause / Remedy

EEPROM FAULTY!

The EEPROM in the recorder is faulty, the configuration
data can no longer be stored.

ADC FAULTY!

The A/D converter of the recorder is faulty.

HALLSENSOR FAULTY!

The writing system is faulty.