beautypg.com

8 programming 67 – JUMO 706021 LOGOLINE 500d Line Recorder Operating Manual User Manual

Page 67

background image

8 Programming

67

Display

Edit

Selection/input

with keys

Enter

Continue with

5

(POTENTIO-
METER)

1 CHARACT.: linear

linear
X1

(customized lin. 1)

X2

(customized lin. 2)

h

,

R

E

customized X1
or X2:

10 (FILTER)

otherwise:

51

51

1 CONNECTION: 2/3-wire

2/3-wire
4-wire

Define type of potentio-
meter connection:
(2/3-wire circuit or 4-wire
circuit)

h

,

R

E

52

52

1 Ro:
+300.0 Ohm

Input start resistance

h

,

R

,

l

,

r

,

S

E

Check
0

Ro

3900 Ohm

53

53

1 Rp:
+900.0 Ohm

Input measurement
range (resistance Rp)

h

,

R

,

l

,

r

,

S

E

Check
0

RANGE

3900 Ohm

Total
resistance
Ro + Rp

3900Ohm

min. span:
6 Ohm

10 (FILTER)

6

see table below!