beautypg.com

Red Lion TSC User Manual

Page 85

background image

CALIBRATION

When calibration is required (generally every two years),

this procedure should only be performed by qualified

technicians using appropriate equipment. Equipment source

accuracies of 0.01% or better are required.

The procedure consists of four parts: applying accurate

mV signals, setting the thermocouple cold junction

temperature, applying precision resistances and measuring

accurate mA currents. Allow a 30 minute warm-up period

before starting this procedure. Do not use thermocouple

wire at any stage of calibration.

This procedure may be aborted by disconnecting power

to the controller before exiting the configuration mode. The

existing calibration settings remain in affect.
Note: After completing any of the calibration sequences, the

controller will default the input sensor type to

thermocouple type “j” (tc-j). Be sure to set input sensor

for proper type.

-83-

Millivolt Calibration (CAL)

Connect precision millivolt source with an accuracy of 0.01% to terminals (+) #9 and (-) #10.

Display

Parameter

Description/Comments

Code

Enter factory

service

48

Calibrate instrument

function code

CAL

Millivolt

yes/no

Calibration required for both RTD and TC input. If this

calibration

procedure is performed the cold junction temp. and RTD

ohms calibration procedures in turn must be completed.

CJC

Thermocouple

yes/no

Not required if only using RTD input. This procedure can

cold junction

only be performed AFTER an accurate mV calibration.

temperature

calibration

rtd

RTD

yes/no

Not required if only using TC input. This procedure can

resistance

only be performed AFTER an accurate mV calibration.

calibration

ANCL

analog output

yes/no

This parameter does not appear if analog output option is

not installed.

Display

Parameter

Description/Comments

StP1

0.0 mV step

Apply 0.0 mV, wait 10 seconds, press PAR

StP2

9.0 mV step

Apply 9.0 mV, wait 10 seconds, press PAR

StP3

18.0 mV step

Apply 18.0 mV, wait 10 seconds, press PAR

StP4

27.0 mV step

Apply 27.0 mV, wait 10 seconds, press PAR

StP5

36.0 mV step

Apply 36.0 mV, wait 10 seconds, press PAR

StP6

45.0 mV step

Apply 45.0 mV, wait 10 seconds, press PAR

StP7

54.0 mV step

Apply 54.0 mV, wait 10 seconds, press PAR

Stp-

Pause

The controller imposes a 5 second delay. (Keep the 54mV signal applied)

The unit advances to CJC - NO.

Configure Step 9 - Factory Service Operations (9-FS)