Chapter 4 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 97
Getting Measurement Message from
Serial Line
Starting Continuous Outputting
R
Enter the R command to start the continuous output of measurements.
The default output format depends on the quantities your transmitter
supports. You can change the output message format and interval with
the FORM and INTV commands. See sections FORM on page 101 and
INTV on page 122.
Example:
>r
P= 1021.6 hPa T= 23.2 'C RH= 5.8 %RH
>
If a value is too long to fit to the allocated space in the output, or if there
is an error in outputting the quantity, the value is displayed with stars '*'.
Example:
RH=***.* %RH T= 31.0 'C
Stopping Continuous Outputting
S
Use the S command to end the RUN mode. After this command all other
commands can be used. You can also press the Esc button or reset the
transmitter to stop the outputting.
See command SMODE to change the default (power-up) operation
mode.
Outputting Reading Once
SEND
Use the SEND command to output the reading once in STOP mode.
Example:
>send
P= 1021.6 hPa T= 23.3 'C RH= 5.7 %RH
>