Chapter 8 _____________________________________________________________ Technical data

VAISALA _______________________________________________________________________ 179

CHAPTER 8

TECHNICAL DATA

This chapter provides the technical data of the product.

Specifications

Performance

Barometric Pressure

Pressure range

500 ... 1100 hPa,

50 ... 1100 hPa

Accuracy

500 ... 1100 hPa,

500 ... 1100 hPa

50 ... 1100 hPa

Class A

Class B

Linearity

±0.05 hPa

±0.10 hPa

±0.20 hPa

Hysteresis*

±0.03 hPa

±0.03 hPa

±0.08 hPa

Repeatability*

±0.03 hPa

±0.03 hPa

±0.08 hPa

Calibration
uncertainty**

±0.07 hPa

±0.15 hPa

±0.20 hPa

Accuracy
at +20 °C***

±0.10 hPa

±0.20 hPa

±0.30 hPa

Temperature
dependence****

±0.1 hPa

±0.1 hPa

±0.3 hPa

Total accuracy

(-40 ... +60 °C/
-40 ... +140 °F)

±0.15 hPa

±0.25 hPa

±0.45 hPa

Long-term
stability/year

±0.1 hPa

±0.1 hPa

±0.2 hPa

Response time
(100 % response)

one sensor

2 s•

1 s•

1 s•

Pressure units

hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia


*

Defined as ±2 standard deviation limits of endpoint non-linearity,
hysteresis error or repeatability error and calibration.

**

Defined as ±2 standard deviation limits of accuracy of the working
standard including traceability to NIST.

***

Defined as the root sum of the squares (RSS) of endpoint non-linearity,

hysteresis error, repeatability error and calibration uncertainty at

room temperature.

****

Defined as ±2 standard deviation limits of temperature dependence
over the operating temperature range.