Data filtering for humidity and temperature, Using display/keypad, Table 32 – Vaisala PTU300 User Manual
Page 115: Filtering levels for humidity temperature

Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 113
Data Filtering for Humidity and
Temperature
The data averaging filter setting for humidity and temperature is separate
from the pressure filtering setting. You can change it using the
display/keypad or on the serial line. There are three levels of filtering
available. The lowest measurement noise is achieved with the extended
filtering.
Table 32
Filtering Levels for Humidity Temperature
Setting
Filtering Level
OFF
No filtering (default setting).
STANDARD
Standard filtering. Approximately 13 s moving average.
EXTENDED
Extended filtering. Approximately 1 min average by
default, but configurable on the serial line.
Using Display/Keypad
1.
Press any of the arrow buttons to open the Main Menu.
2.
Select Measuring
by pressing the ► arrow button.
3.
Select Filtering and press CHANGE to confirm your selection.
4.
Select Off/Standard/Extended and press SELECT to confirm
your selection.
5.
Press EXIT to return to the basic display.