Data filtering for pressure – Vaisala PTU300 User Manual
Page 114

User's Guide _______________________________________________________________________
112 __________________________________________________________________ M210796EN-G
Data Filtering for Pressure
Use the serial command AVRG to view and set the averaging time
during which the individual pressure measurement samples are integrated
to get an averaged reading. The averaging time is the total averaging time
of the transmitter.
AVRG [x]
where
X
=
1 ... 60 s (default: 1 s)
.
Note that if the averaging time is defined to be long, the settling time at
power-up will be long, too.
A minimum of one-second averaging time is recommended per each
pressure transducer. These selections are used as the factory setting
averaging times.
Example of setting the averaging time to 60 seconds (WMO averaging
time for barometric pressure measurement):
>avrg
P1 average : 1 s ? 60
>