Using Serial Line

PRES and XPRES

Command XPRES must be used if the value is changed frequently (for
example, by an automatic system that updates the value). Its value is not
retained at reset, and when set to 0, last value set with PRES is used
instead. Use the serial line and do the following:

PRES [aaaa.a]

XPRES [aaaa.a]

where

aaaa.a = Absolute process pressure (hPa)

Example:

>pres
Pressure : 1013.00 hPa ?
>pres 1010
Pressure : 1010.00 hPa
>

Table 31

Conversion Factors for Pressure Units

From

To: hPa

mbar

1

Pa N/m2

0.01

mmHg torr

1.333224

inHg

33.86388

mmH

2

O

0.09806650

inH

2

O

2.490889

atm

1013.25

at

980.665

bar

1000

psia

1)

68.94757

1) psia = psi absolute.

Example:

29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa

PFIX

Use the PFIX command to select either P inputted or P measured.

- When PFIX is On, fixed PRES value is used

- When PFIX is Off, measured PRES value is used