User's Guide _______________________________________________________________________

158 __________________________________________________________________ M210796EN-G

Table 36

Error Messages

Error Code Error Message

Action

E0

Humidity sensor measurement
malfunction.

Check the integrity of the humidity probe
and the probe cable. Clean the probe from
dirt, water, ice or other contaminants.

E1

Humidity sensor short circuit

E2

Humidity sensor open circuit

Check the integrity of the humidity probe
and the probe cable.

E3

Temperature sensor open circuit.

E4

Temperature sensor short circuit. Check the integrity of the humidity probe

and the probe cable. Clean the probe from
dirt water, ice or other contaminants.

E5

Temperature measurement
malfunction

E6

Temperature sensor current leak. Check the integrity of the humidity probe

and the probe cables. Clean the probes
from dirt, water, ice or other contaminants.

E7

Internal ADC read error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E8

Additional temperature sensor
short circuit

Check the integrity of the temperature
probe and the probe cable. Clean the
probe cable from dirt, water, ice or other
contaminants.

E9

Checksum error in the internal
configuration memory

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E10

Internal EEPROM read error

E11

Internal EEPROM write error

E12 … E13 Add-on module 1 (or 2)

connection failure

Turn off the power and check the module
connection. Turn on the power.

E14

Device internal temperature out of
range

Ensure that the operating temperature is
within the valid range.

E15

Internal RAM memory or software
checksum error

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E16 … E17 Pressure measurement failure in

add-on module slot 1or 2

Disconnect power and check pressure
module connection.

E18

Internal ADC reference voltage
out of range

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E19

Internal analog output reference
voltage out of range

E20 … E22 Configuration switches for analog

output 1/2/3 set incorrectly

Check and re-set the switches, see page
55.

E24 … E25 Internal error in add-on module

1 (or 2)

Disconnect the power and check the
module connection.

E26

Communication module installed
in incorrect add-on module slot

Disconnect the power and change the
communication module to another module
slot.

E27

Pressure out of valid range

Check that assumed pressure is within
measurement range for the transmitter.

E28 … E29 Unknown/incompatible module

installed in add-on module slot 1
(or 2)

Ensure that the module is compatible with
the PTU300.

E30

Internal analog voltage out of
range

Internal transmitter failure. Remove the
transmitter and return the faulty unit to
Vaisala Service.

E31

Internal system voltage out of
range

Check that the supply voltage is correct
and the power supply can provide enough
power for the instrument.