Device information, Using serial line – Vaisala PTU300 User Manual
Page 117

Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 115
Device Information
The device information contains current configuration; status and settings
of the device. The information is available through the display/menu, as
well. For more information, see Information Display on page 72.
When requesting the device information, the following information will
be shown:
- Current sensor operation (for example, chemical purge), if any, in
progress
- Present or past unacknowledged errors, if any
- Device identification; product name, version and serial number
- Current date and time (only shown if data logger module installed)
- Adjustment information (the latest date, user made adjustments)
- Measuring settings
- Information on chemical purge settings (when applicable)
- Display Alarm settings
- Serial interface information
- Network settings and status of the LAN and WLAN interfaces
- Analog output information
- Relay output information (when applicable)
Using Serial Line
?
Use the serial line command ? to check the current transmitter
configuration. Command ?? is similar but can also be used if the
transmitter is in POLL mode.
Example:
>?
PTU300 / 3.01
Serial number : A2150004
Batch number : A1450004
Adjust. date : 2006-01-22
Adjust. info : (not set)
Date : 2006-02-27
Time : 14:00:57
Serial mode : RUN
Baud P D S : 4800 E 7 1
Output interval: 150 s
Address : 0
Echo : OFF
Pressure : 1013.25 hPa
Filter : ON