beautypg.com

Analog output tests, Itest, On analog output tests on – Vaisala PTU300 User Manual

Page 132

background image

User's Guide _______________________________________________________________________

130 __________________________________________________________________ M210796EN-G

Analog Output Tests

Use the display/keypad to test the operation of the analog outputs by
forcing the outputs to known values. Measure then the outputs with a
current/voltage meter.

1.

Press any of the arrow buttons to open the Main Menu.

2.

Select System and press the right arrow button.

3.

Select Diagnostics and press the right arrow button.

4.

Select Analog output tests and press the right arrow button.

5.

Select one of the testing options Force 0%/50%/100% of scale.
Press the TEST key. All outputs are tested simultaneously. The
actual output value depends on the selected range.

6.

Press the OK key to stop testing. Then press the EXIT key to
return to the basic display.

ITEST

Use the serial line to test the operation of the analog outputs. Use the
ITEST command to force the analog outputs to entered values. The set
values remain valid until you enter the command ITEST without
parameters or RESET the transmitter.

ITEST [aa.aaa bb.bbb cc.ccc]

where

aa.aaa = Current or voltage value to be set for channel 1 (mA or V)
bb.bbb = Current or voltage value to be set for channel 2 (mA or V)
cc.ccc = Current or voltage value to be set for channel 3 (optional)

(mA or V)

Example:

>itest 20 5
Ch1 (Td ) : * 20.000 mA H'672A
Ch2 (T ) : * 5.000 mA H'34F9
>itest
Ch1 (Td ) : -23.204 'C 16.238 mA H'FFFE
Ch2 (T ) : 22.889 'C 8.573 mA H'5950
>