Measurement data (read-only), Table 45, Measurement data registers – Vaisala PTU300 User Manual
Page 212: Table 45 measurement data registers

User's Guide _______________________________________________________________________
210 __________________________________________________________________ M210796EN-G
integer range. Maximum writable value is 32767 also for configuration
registers that do not accept negative values.
Measurement Data (Read-Only)
Table 45
Measurement Data Registers
Name
Float
Integer
Unit
RH
0001…0002
0257 (×0.01)
%
T
0003…0004
0258 (×0.01)
°C
T
a
0005…0005
0259 (×0.01)
°C
T
d
0007…0008
0260 (×0.01)
°C
T
d/f
0009…0010
0261 (×0.01)
°C
a
0015…0016
0264 (×0.01)
g/m³
x
0017…0018
0265 (×0.01)
g/kg
T
w
0019…0020
0266 (×0.01)
°C
H
2
O
0021…0022
0267 (×1)
ppm
V
p
w
0023…0024
0268 (×0.1)
hPa
p
ws
0025…0026
0269 (×0.1)
hPa
H
0027…0028
0270 (×0.01)
kJ/kg
ΔT
0031…0032
0272 (×0.01)
°C
P
0043…0044
0278 (×0.01)
hPa
QNH
0045…0046
0279 (×0.01)
hPa
QFE
0047…0048
0280 (×0.01)
hPa
HCP
0049…0050
0281 (×0.01)
hPa
P
3h
0051…0052
0282 (×0.01)
hPa
P
1
0053…0054
0283 (×0.01)
hPa
P
2
0055…0056
0284 (×0.01)
hPa
H
2
O
0065…0066
0289 (×1)
ppm
W
P
3h
WMO tendency code
0067…0068
0290 (×1)
Available measurements depend on the instrument configuration. Values
may be unavailable also in case of device failure. Read status registers or
exception status outputs to check for failures.
NOTE
Only real-time data can be read from the instrument. Data that is
recorded in the transmitter’s memory (internal memory or data logger
module) cannot be read through Modbus.