Relative Humidity Adjustment After Sensor
Change

Using Display/Keypad

When using the optional display/keypad, follow the instructions in
section Using Display/Keypad on page 168 on page but select Adj. for
new RH sensor (
instead of 1-point/ 2-point adjustment).

Using Serial Line

After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.

FCRH

Example:

>fcrh
RH

:

1.82 1. ref

?

0

Press any key when ready...

RH

:

74.22

2. ref

? 75

OK
>

The

OK

indicates that the calibration has succeeded.