Modbus, Overview of modbus protocol support, Chapter 5 – Vaisala PTU300 User Manual
Page 149: Table 35, Supported modbus variants
Chapter 5 ___________________________________________________________________ Modbus
VAISALA _______________________________________________________________________ 147
CHAPTER 5
MODBUS
This chapter contains information that is needed when operating the
transmitter using the Modbus protocol.
Overview of Modbus Protocol Support
The PTU300 transmitter can be accessed using the Modbus serial
communication protocol. Support for Modbus protocol is available on all
PTU300 transmitters as a standard feature from software version 5.10
onward. The supported Modbus variants and the connections they use are
listed in Table 35 below.
Table 35
Supported Modbus Variants
Supported Modbus Variant
Connections
Modbus RTU (Serial Modbus)
RS-232 (standard User Port)
RS-422/485 interface (optional module)
Modbus TCP (Ethernet Modbus)
LAN interface (optional module)
WLAN interface (optional module)
The supported Modbus functions, registers, configuration options, and
diagnostics are described in Appendix C, Modbus Reference, on page
207.
Note the following limitations of the Modbus implementation:
- Modbus TCP mode accepts only one TCP connection at a time.
Design the system so that only one Modbus TCP client accesses the
transmitter.
- Modbus TCP can process reliably only one Modbus transaction at a
time. Reduce the polling rate of the client to avoid nested transactions.