User's Guide _______________________________________________________________________

102 __________________________________________________________________ M210796EN-G

where

x

=

Formatter string

The formatter string consists of quantities and modifiers. If no formatter
string is entered, the command will display the currently active formatter
string. Note that the hash symbol "#" is shown as a backslash "\" when
viewing the current formatter string.

When entering the command, use the abbreviations of the quantities. For
more information on quantities, see Table 4 on page 22, Table 5 on page
22, and Table 6 on page 22.

The modifiers are presented in Table 30 on page 102.

Table 30

FORM Command Modifiers

Modifier

Description

x.y

Length modifier (number of digits and decimal places)

#t

Tabulator

#r

Carriage-return

#n

Line feed

“”

String constant

#xxx

Special character, code "xxx" (decimal), for example #027 for
ESC

U5

Unit field and length (length optional)

ADDR

Transmitter address [00...255]

ERR

Error flags for P, T, Ta, RH [0000 ... 1111], 0 = no error

STAT

Transmitter status in 7 character field, for example:
N 0

no heating

h 115

probe heating active, power 115/255

H 159.0

purge heating active, temperature 159ºC

S 115.0

purge cooling active, temperature 115ºC

X 95.0

sensor heating active, temperature 95ºC

SN

Transmitter serial number

TIME

Time [hh:mm:ss]

DATE

Date [yyyy-mm-dd]

OK

Pressure stability indicator, two characters [OK or " "]

CS2

Modulus-256 checksum of message sent so far, ASCII encoded
hexadecimal notation

CS4

Modulus-65536 checksum of message sent so far, ASCII
encoded hexadecimal notation

CSX

NMEA xor-checksum of message sent so far, ASCII encoded
hexadecimal notation

A3H

Pressure tendency [* or 0...8]