Taking modbus into use – Vaisala PTU300 User Manual
Page 150

User's Guide _______________________________________________________________________
148 __________________________________________________________________ M210796EN-G
Taking Modbus into Use
To take the Modbus protocol into use on the PTU300, you must perform
some configuration tasks using the built-in display and keypad (optional)
or a PC connected to the serial line. For example, you can connect to the
service port using the USB service cable (Vaisala order code: 219685).
The transmitter must be powered from a suitable power supply during
configuration.
Instructions for performing the configuration are provided in the
following sections:
- Enabling Serial Modbus on page 149
- Enabling Ethernet Modbus on page 150
After configuring the transmitter, perform the transmitter installation and
wiring according to the instructions in Chapter 3, Installation, on page
29.
The configuration instructions assume you are familiar with the use of
the transmitter. Additional details on the display/keypad, service port use,
and serial commands are available in other sections of this manual:
- The use of the display/keypad option is described in section
Display/Keypad (Optional) on page 66.
- If this is the first time you are using the USB service cable, you can
find detailed instructions in section Service Port Connection on page
82. You must install the required driver before using the service cable.
- Configuration of the LAN and WLAN interfaces is decribed in section
- The serial commands available through the service port are described
beginning from section List of Serial Commands on page 94.