beautypg.com

Inputs and outputs – Vaisala PTU300 User Manual

Page 187

background image

Chapter 8 _____________________________________________________________ Technical data

VAISALA _______________________________________________________________________ 185

Inputs and Outputs

Operating voltage

10 ... 35 VDC, 24 VAC

with optional power supply module

100 ... 240 VAC, 50/60 Hz

Start-up time after power-up

3 s


Power consumption at 20 °C (Uin 24VDC)

RS-232

max 28 mA

Uout 3 x 0...1V / 0...5V / 0...10V

max 33 mA

Iout 3 x 0...20 mA

max 63 mA

display and backlight

+ 20 mA

during chemical purge

+ 110 mA max

during probe heating (PTU307)

+ 120 mA max

Settling time at power-up (one sensor)

class A

4 s

class B

3 s

Analog outputs

current output

0 ... 20 mA, 4 ... 20 mA

voltage output

0 ... 1 V, 0 ... 5 V, 0 ... 10 V

Humidity and temperature

Accuracy of analog outputs at 20 °C

± 0.05 % full scale

Temperature dependency of the

analog outputs

± 0.005 %/°C full scale

Pressure

500...1100 hPa 50...1100 hPa

Accuracy of analog outputs at 20 °C

0.30 hPa

0.4 hPa

Temperature dependency of the

analog outputs

0.6 hPa

0.75 hPa

External loads

current ouputs

RL < 500 ohm

0... 1V output

RL > 2 kohm

0... 5V and 0... 10V outputs

RL > 10 kohm

Wire size

0.5 ... 2.5 mm

2

(AWG 20 ... 14)

stranded wires recommended

Digital outputs

RS-232

RS-422/485 (optional)

LAN (optional)

WLAN (optional)

Protocols

ASCII commands

Modbus RTU

Modbus TCP

Relay outputs (optional)

0.5 A, 250 VAC, SPDT

Display (optional)

LCD with backlight, graphic
trend display

Menu languages

Chinese, English, Finnish,
French, German, Japanese,
Russian, Spanish, Swedish