Help, Errs, Mods – Vaisala PTU300 User Manual
Page 119

Chapter 4 _________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 117
HELP
Use the HELP command to list the commands.
Example:
>help
? ACAL ADDR AERR ALSEL
ASEL AVRG CDATE CLOSE CON
CRH CT CTEXT DATE DELETE
DIR DSEL DSEND ECHO ERRS
FCRH FILT FORM HELP HHCP
HQFE HQNH INTV ITEST LIGHT
LOCK MODBUS MODS MPC MPCI
OFFSET PDMAX PFIX PLAY PRES
PSTAB PUR PURGE R RESET
SCOM SDELAY SEND SERI SMODE
SWARM SYSTEM TEST TIME UNDELETE
UNIT VERS XHEAT XPRES
>
ERRS
Use the ERRS command to display transmitter error messages, see Error
States on page 157 and Table 36 on page 158.
Example:
>errs
No errors
>
Example:
>errs
Error: E2 Humidity sensor open circuit.
>
MODS
Use the MODS command to display information about the optional
modules that are connected to the transmitter.
Example:
>mods
Module 1 : WLAN-1 (POST: release_82000941_J; FW:
Version 82000977_K1 10/16/2008)
Module 2 : LOGGER-1 (1024 MB; HW: B; SW: 5)
>