beautypg.com

Configuring holdoff – GW Instek GDS-1000A-U Series User Manual User Manual

Page 98

background image

GDS-1000A-U Series User Manual

98

line

1~263 for NTSC, 1~313 for PAL/SECAM

Pulse condition
(pulse trigger)

Sets the pulse width (20ns ~ 10s) and the triggering
condition.

>

Longer than =

Equal to

<

Shorter than ≠

Not equal to

Trigger slope

Triggers on the rising edge.

Triggers on the falling edge.

Trigger coupling AC

Triggers only on AC component.

DC

Triggers on AC+DC component.

Frequency
rejection

LF

Puts a high-pass filter and rejects the
frequency below 50kHz.

HF

Puts a low-pass filter and rejects the
frequency above 50kHz.

Noise rejection

Rejects noise signals.

Trigger level

LEVEL

Using the trigger level knob moves the
trigger point up or down.

Configuring Holdoff

Background

The Holdoff function defines the waiting period
before GDS-1000A-U starts triggering again after
the trigger point. The holdoff function is especially
useful for waveforms with two or more repetitive
frequencies or periods that can be triggered.

Panel operation

1. Press the Trigger menu key

twice.

MENU

MENU

2. To set the Holdoff time, use

the Variable knob. The
resolution depends on the
horizontal scale.

Range

40ns~2.5s