8.3 Storage Space in the Display Memory
The total space available in the Display Memory to store data is
131072 measurement cycles. Depending on the time for one
measurement cycle the capacity will be as follows.
If a measurement cycle in which gas A and water vapour are
measured with continuous sampling takes 15s, then the Display
memory space will be 546 hours = 22 days.
If a measurement cycle where 5 gases and water vapour are
measured with continuous sampling takes 60 s, then the Display
memory space will be 2184 hours = 91 days.
The Monitor will overwrite the oldest stored measurement data when
the Display memory is full. This is done by deleting data in blocks of
2048 measurement cycles.
For further information regarding measurement cycles see