3 checking/changing the text line terminator, 6 setting the ethernet communication parameters, Section 2.6 – LumaSense Technologies INNOVA 1314i User Manual
Page 18
![background image](/manuals/363951/18/background.png)
Chapter 2
______________________________________________________________________
_____________________________________________________________________________
BE6030-13
1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 18 of 199
If the baud rate displayed is incorrect press
¿
and use ▲ and ▼ to
display the correct value. Press
¿
again to store the selection.
If the baud rate displayed is correct, then press ▼ to continue to the
next parameter.
Press S1 to select 1 STOP BIT.
Press S1 to select 7 DATA BITS
Press S2 to select EVEN PARITY
Press S3 to select LEASED-LINE
Press S3 to select HARD-WIRED HANDSHAKE.
Press SET-UP to exit the set-up mode
Press RESET and S1 in order that the new settings are enabled.
The monitor and the Gas Monitoring Software 7304 are now able to
communicate together.
2.5.3
Checking/Changing the Text line Terminator.
The Text line Terminator must be set using the push-buttons on the
front of the monitor.
Press SET-UP S2 ▼S3
Press SET-UP to exit the set-up mode
2.5.4
Checking/Changing the Data Log and Error Log
Options.
The Data Log and Error Log must be set using the push-buttons on
the front of the monitor.
Press SET-UP S3 S1 S1 S3 ▼S1 S1
Press SET-UP to exit the set-up mode
2.6 Setting the Ethernet Communication Parameters
The LumaSoft Gas 7810 application with Ethernet communication is
able to communicate with the 1314i gas monitor through a local
Ethernet network.
The homepage of the 1314i gas monitor can also be displayed in a
web browser, when the 1314i gas monitor is connected to a local
Ethernet network.