1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 197 of 199
X-ON/X-OFF HANDSHAKE CONFLICT
The Monitor receives the
handshake signal from the
RS232 CTS ( clear to send)
line, but it has not been
set-up to use this CTS line
The Monitor and
RS232 device
(printer) do not
agree about which
handshake signal
they should use.
Perhaps the wrong
interface cable has
been used.
TIMEOUT. DEVICE NOT CONNECTED
Monitor wishes to send
data (data-log or
error-log) via the IEEE
interface, but there is no
response from the device
with the address entered
in the Monitor set-up
1 No device is con-
nected to the
Monitor
2 The device
address has
been entered in-
correctly in the
Monitor’s set-up.
TIMEOUT. DEVICE NOT READY
Monitor wishes to send
data
(print, data-log or error-
log) but after the start of
data transmission, the
device stops receiving
data.
1. The device is
“off-line“(it
cannot receive
data).
2. The device is not
able to receive
data fast
enough.
3. The device is no
longer
connected to
Monitor.