2 the measurement parameters, Sampling – LumaSense Technologies INNOVA 1314i User Manual
Page 77
Chapter 7
______________________________________________________________________
_____________________________________________________________________________
BE6030-13
1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 77 of 199
Units branch
2.
Press S2 to enter the Units branch.
3.
Use the select push-buttons to define the concentration unit.
4.
Use the select push-buttons to define the length unit.
5.
Use the select push-buttons to define the temperature unit.
6.
Use the select push-buttons to define the pressure unit.
7.
The humidity unit is displayed. Use the direction push-buttons
to change it to the correct value.
When step 7 is completed the display returns to the top of the Units
branch.
Press p to go to the top of the Set-Up Tree, or press SET-UP to exit
the set-up mode.
7.2 The Measurement Parameters
Sampling
Sampling mode: the manner in which the gases are sampled. There
are two possibilities here:
Continuous sampling: once the monitor is finished measuring it
starts again making the next measurement,
Interval Sampling:
this is the time from the start of one
measurement cycle to the start of the next
one, for a full explanation see
SELECT CONFIGURATION SET-UP BRANCH
SYSTEM UNIT FILTERS