1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 143 of 199

GAS A IS “_ “

PRESS ENTER TO CHANGE VALUE

7. Press

¿

and use the direction push-buttons to enter the desired

name.


8. Press

¿

to “accept” the name you have “entered” on the display

screen. The following text then appears:

MOLECULAR WEIGHT OF GAS A 1.000

PRESS ENTER TO CHANGE VALUE


9. Press

¿

and use the direction push-buttons to “enter” the

molecular weight of the gas to be measured by filter A.

10. Press

¿

to “accept” the molecular weight you have “entered” on

the display screen. The following text appears on the screen:

FILTER A SAMPLE INTEGRATION TIME 5 s

PRESS ENTER TO CHANGE VALUE

11. Press

¿

, and use p and q to display the desired Sample

Integration Time.

There are seven to choose from: 0.5s, 1s, 2s, 5s, 10s, 20s and

50s.

12. Press

¿

to “accept” the selected value.

The following text appears on the screen:

GAS A HIGH ALARM LIMIT 1 ___________mg/m

3

PRESS ENTER TO CHANGE VALUE

13. Press

¿

and use the direction push-buttons to enter the desired

alarm level.

14. Press

¿

to accept the entered value. Following screen is

displayed:

GAS A HIGH ALARM LIMIT 2 ___________mg/m

3

PRESS ENTER TO CHANGE VALUE

15. Repeat step 13 and 14 to enter the desired alarm level.