LumaSense Technologies INNOVA 1314i User Manual
Page 196
![background image](/manuals/363951/196/background.png)
Chapter 15
______________________________________________________________________
_____________________________________________________________________________
BE6030-12
1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 196 of 199
CLOCK SET TO DEFAULT ERROR
DETECTED IN CLOCK SETTINGS
The Monitor’s internal
clock was found to be
incorrectly set so these
values have been set to
their default values
A new battery has
been installed
FACTORS SET TO DEFAULT ERROR
DETECTED IN CALIBRATION
FACTORS
An error has been found in
the data stored in the
Calibration Factor part of
the Monitor’s memory, so
these factors have been
set to their default values
1. A defective
back-up battery
2. A defective
Working Memory
(RAM) or Source
Memory
(EEPROM)
3. A Software fault
MEMORY SET TO DEFAULT ERROR
DETECTED IN BACKGROUND MEMORY
An error has been detected
in the Monitor’s
Background Memory, so
the Monitor automatically
deletes all data stored in
this Memory
1. A defective
back-up battery
2. A defective
Working
memory ( RAM )
3. A software fault
MEMORY SET TO DEFAULT ERROR
DETECTED IN DISPLAY MEMORY
An error has been detected
in the Monitor’s Display
Memory, so the Monitor
automatically deletes all
data stored in this
memory.
1. A defective
back-up battery
2. A defective
Working Memory
( RAM )
3. A software fault
MEMORY SET TO DEFAULT ERROR
DETECTED IN INTERNAL MEMORY
An error has been found in
the data stored in the
Monitor’s internal memory.
data stored in this memory
cannot be read or altered
by the user, so the Monitor
automatically corrects any
errors found in this
memory
1. A defective
back-up battery
2. A defective
Working Memory
( RAM)
3. A software fault.
MEMORY SET TO DEFAULT ERROR
DETECTED IN SOURCE MEMORY
An error has been found in
the data stored in the
Monitor’s source memory,
so the Monitor
automatically gives stored
parameters default values.
1. A defective
back-up battery
2. A defective
Source Memory
( EEPROM)
3. A software fault
PARAMETERS SET TO DEFAULT
ERROR DETECTED IN
CONFIG:/FORMAT PARAMETER
An error has been found in
the value parameters
stored in the Configuration
and/or Format branches of
the Monitor’s set-up tree,
and therefore the Monitor
has automatically given
these parameters default
values
1. A defective
back-up battery.
2. A defective
Working Memory
(RAM)
3. A software fault.
PARAMETERS SET TO DEFAULT
ERROR DETECTED IN THE TASK
SET_UP PARAMETER
An error has been found in
the value parameter stored
in the Monitoring Task
branch of the Monitor’s
set-up tree, and has
automatically given these
parameters default values.
1. A defective
back-up battery
2. A defective
Working Memory
(RAM)
3. A software fault.