1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 96 of 199
“W”
This symbol indicates that a warning was detected
during this measurement. The accuracy of such a
measurement cannot always be guaranteed.
Gas Marks:
“B”
The Monitor was incapable of calculating this gas
concentration.
“F”
After this measurement the filter carousel was
found to be out of alignment. If the carousel was
only slightly out of alignment then the accuracy of
the measurement has not been affected. The ac-
curacy of such a measurement cannot always be
guaranteed.
“A”
The gas concentration measured during this
measurement cycle was found to exceed the user-
defined alarm limit for this gas.
8.2.3
Symbols Used in Print-outs of Measurement Data
Displayed measurements marked with the COMMON MARKS “O”
and “W”, and the GAS MARKS “A” and ”F”, are printed-out with an
“X” symbol. However, any displayed measurements marked with the
COMMON MARKS “P” and the GAS MARK “B” are not printed-out
with these symbols. These measurements are marked in a different
way:
Common Marks:
The symbol “P” is replaced by the text “1314i Restarted”. This
indicates that the Monitor stopped and then started-up again
sometime between the previous sample time and the time which
appears alongside this message.
Gas Marks:
The symbol “B” is replaced by a straight line (_______) which
indicates that the Monitor was unable to calculate the concentration
of the named gas in this sample.
Other Marks Used:
……This symbol indicates either: (1) that this gas was not measured;
or (2) that this gas was not selected to be printed out (see