beautypg.com

LumaSense Technologies INNOVA 1314i User Manual

Page 189

background image

Chapter 14

______________________________________________________________________

_____________________________________________________________________________

BE6030-12

1314i Photoacoustic Gas Monitor

LumaSense Technologies A/S

Page 189 of 199

Fig.14.1 Curve illustrating the two points measured during a two-

point span calibration task

GAS X: LOW MEAS. > HIGH MEAS.

This message indicates that the signal measured in the cell when the

gas concentration was low is found to be greater than the signal when

the gas concentration was high.

The above message could indicate that the span gas concentrations

were incorrectly “entered” when the task was set-up (that is, the

“entered” low gas concentration was actually greater than the

“entered” high gas concentration). Alternatively, it could indicate that

the concentration of the gas supplied to the Monitor during the

measurement of V

low

was actually the high concentration (C

high

) and

vice versa.

The signals V

high

and V

low

have to be corrected for water-vapour

interference and cell noise, that is:

V

high

– V

offset

– V

H2O

= V

h

;

and:

V

low

– V

offset

– V

H2O

= V

l

These two points (V

h

, C

h

) and V

l

, C

l

) can then be plotted on a graph

(see

Fig.14.1

).

When the two signals have been measured, the Monitor then finds the

value δ = 1% of V

h

and finds out the relationship between the

measured values: