1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 83 of 199
Environment branch
13. Press S3 to enter the Environment branch.
14. Use the select push-buttons to select the desired flushing
mode.
If you select Auto, go to step 17.
If you select Fixed Time, go to step 15.
15. Use the direction push-buttons to define the Chamber Flushing
Time.
16. Use the select push-buttons to select Tube Flushing.
If you don’t want tube flushing, select NO and go to step 18.
If you select tube flushing, use the direction push-buttons to
define the flushing time and go to step 18.
17. Use the direction push-buttons to select the desired sample
tube length and move on to the next display.
18. Use the direction push-buttons to select the desired
normalization temperature.
for details about the normalization options.
When step 18 is completed, the display returns to the top of the
MEASUREMENT branch.
Press p to go to the top of the Set-up Tree.
Press SET-UP to exit the set-up function.
Format branch
If you are starting here, press SET-UP:
19. Press S2 to enter the FORMAT branch.
20. Use the select push-buttons to choose if you want to print each
gas.
If you select Yes, go to step 22.
If you select No, go to step 23.
21. Use the select push-buttons to choose which individual
parameters should be printed. Each time one of the select
push-buttons is pressed, the next parameter is displayed. This
step can be repeated up to 7 times, depending on the number
of filters installed.
22. Use the select push-buttons to select the Text line Terminator.