beautypg.com

Yokogawa Button Operated MV2000 User Manual

Page 94

background image

3-20

M-4660

3.11 Setting Range-Out Detection for Linearly

Scaled Measurement Channels

Display

RD-MV1000

Press MENU > then select Menu tab > Basic setting mode > Environment tab > Input,
Alarm

RD-MV2000

Press MENU and then select Menu tab > Basic setting mode > Environment tab >
View, Message, Input, Alarm.

Settings

• Input > Value on over-range

Setting

Description

Free

Ignoring the decimal point, less than –30000 is a negative range-out, and above
30000 is a positive range-out. They are displayed as –Over and +Over.

Over

A value less than –5% of the scale is a negative range-out, and a value greater than
105% of the scale is a positive range-out. They are displayed as –Over and +Over.
Example: If the scale is 0.0 to 200.0, a value less than –10.0 is a negative range-

out, and a value greater than 210.0 is a positive range-out.

Note

With computation and report functions such as TLOG and CLOG, you can specify how the MV
will handle scale range out values.

See section 10.1.

This manual is related to the following products: