• Tag, Memory sample, and Alarm delay

Press MENU and then select Menu tab > Ext. channel > Tag, Memory sample, Alarm

delay.

• Channels to Be Manually Sampled

See section 4.5.

Settings

• Configuring the Input

The measured values of external devices are loaded into external input channels by

using the Modbus client or Modbus master function.

For configuration instructions, see the Communication Interface User’s Manual (IM

RD-MV1000-17E).

• First-CH, Last-CH

Select the target channels. Channel numbers are from 201 to 440.

• Ext. range > On/Off

Select On to use the channels.

• Ext. range > Span Lower, Span Upper

The measurement range.
Selectable range of values: –30000 to 30000
Decimal place: Up to four fractional digits

• Ext. range > Unit

Set the unit (up to six characters,

A a # 1

).

• Ext. alarm

The available alarm types are high limit alarm, low limit alarm, delay high limit alarm,

and delay low limit alarm.

The range of alarm values is as follows:

Type

Alarm Values

Example of Alarm Values

H, L

In the range of –30000 to 30000 ignoring
the decimal point

If the span is 0.0 to 100.0:
–3000.0 to 3000.0

T, t

Same as H, L

Same as H, L

For alarm configuration instructions, see section 3.7.

* If you change the On/Off or span settings of an external input channel, the alarm on that

channel will be set to Off.

11.1 Configuring External Input Channels