9 saving and loading setup data – Yokogawa Button Operated MV2000 User Manual
Page 113
4-19
M-4660
Measurement and Recording
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
4.9
Saving and Loading Setup Data
This section explains how to save and load setup data from external storage media.
Procedure
• Saving Setup Data
1.
To access the display:
Press MENU and then select File tab > Save settings. Press the CF or USB soft
key
*
and then DISP/ENTER.
* When using a CF card and USB flash memory.
2.
Enter the file name (up to 32 characters,
A a # 1
).
You can use these symbols: # % ( ) + – . @ ° _
You cannot use these character combinations: “AUX” “CON” “PRN” “NUL”
“CLOCK” “COM1” to “COM9” and “LPT1” to “LPT9”.
To cancel the operation, press ESC.
3.
Press DISP/ENTER.
The setup data is saved.
• Loading Setup Data
The MV loads all Setting Mode and Basic Setting Mode settings. When setup data is
loaded, the previous settings are erased and the MV restarts with the loaded settings.
1.
To access the display:
*
Press MENU and then select File tab > Load settings. Press the CF or USB soft
key
*
and then DISP/ENTER.
* When using a CF card and USB flash memory.
Note